Art: 714376-00M Rev. Date: 31-Jan-12 14-9
CONTROLS FOR ACT CARTRIDGES
Intended Use
The i-STAT
®
ACT Control Level 1 and ACT Control Level 2 are intended for use
toverifytheintegrityofnewlyreceivedi-STATACTcartridges.Thecontrols
produceclottingtimesexpectedformoderateandhighlevelheparinization
to indicate that the cartridges are functioning properly.
Contents
Eachlevelofcontrolispackagedasaboxof5vialsoflyophilizedhumanplasma
and5vialsof9.5
±
1.5mmol/Lcalciumchloridediluent.
Storage
i-STAT ACTcontrols,Levels1 and2,are containedin6-mLvials. Separate
6-mL vials contain 1-3 mL of calcium chloride solution for reconstitution.
Refrigeratedstorageat2to8ºC(35to46ºF)shouldbemaintaineduntilthe
printedexpirationdateontheboxandviallabels.Donotusebeyondthe
expirationdateontheboxandviallabels.
Control solutions may also be stored at room temperature for up to 4 hours (18
to30°Cor64to
-
86°F).Ifleftoutlongerthan4hoursatroomtemperature,
they should be discarded.
Warnings and
Precautions
Handle this product using the same safety precautions used when handling
any potentially infectious material. The human plasma used in the
preparation of this product has been tested by FDA approved test methods and
foundnegative/non-reactiveforHIV-1,HIV-2,HBsAg,andHCV.However,no
known test method can offer complete assurance that products derived from
humanbloodwillnottransmitinfectiousdisease.
Disposeofthisproductasbiohazardouswasteaccordingtoalllocal,state,and
national regulations.
Directions for Use
Priortotesting,vialscontainingthelyophilizedplasmaandCaCl
2
reconstituting
uidshouldstandatroomtemperature(18-30ºCor64-86ºF)foraminimum
of45minutes.Forbestresults,vials,cartridges,andanalyzersshouldbeatthe
same temperature.
Reconstituteonlyonelevelofcontrolplasmaatatime.CONTROLSOLUTIONS
MUSTBEUSEDIMMEDIATELY(lessthan30seconds)AFTERCOMPLETING
THERECONSTITUTIONANDMIXINGSTEPS.
ii. Note:thisproductisabufferedaqueoussolutioncontaining
electrolytes and non-conductive ingredients. The only value
assigned to this fluid is for hematocrit. All other cartridge analyte
resultsobtainedwiththiscontrolmaterialshouldbeignored.
5. Ifavailable,transmitresultstotheCentralDataStation.