14-14 Art: 714376-00M Rev. Date: 31-Jan-12
i-STAT BNP CONTROLS (LEVELS 1, 2, AND 3)
Intended Use
i-STATBNPControlsareassayedliquidplasmausedtoverifytheintegrityof
newlyreceivedi-STATBNPcartridges.
Warning and
Precautions
Handletheproductsusingthesamesafetyprecautionsusedwhenhandlingany
potentially infectious material. The human plasma used in the preparation of
theseproductshasbeentestedbyFDAapprovedtestmethodsandfoundnegative/
non-reactiveforHIV-I,HIV-2,HBsAg,andHCV.However,noknowntestmethod
canoffercompleteassurancethatproductsderivedfromhumanbloodwillnot
transmit infectious disease.
Each1.0mLvialcontains<0.1%sodiumazideasapreservative.Disposeofthis
productaccordingtolocal,state,andnationalregulations.Ifdisposeddowna
drain,thesodiumazideinthisproductmayreactwithleadandcopperplumbing
toformhighlyexplosivemetalazides.Flushdrainwithlargeamountsofwater
to prevent azide build-up.
Donotuseifthecontrolsarrivethawedoruncapped.
Bacterial contamination of the control can cause an increase in turbidity. Do not
usethecontrolmaterialifthereisvisibleevidenceofmicrobialgrowthorgross
contamination.
Storage and Stability
Store at ≤ -18˚C(-1˚F)inanon-defrostingfreezer.Afterthawing,theopened
orunopenedvialisstablefor4hourswhencappedandstoredat2-8˚C(35-
46˚F).Donotre-freeze.
Procedure
1. Remove vial from freezer and thaw at room temperature (18-30˚C)
for 15 minutes.
2.
Gently invert the vial 10 times, then swirl the vial 10 times. Inspect
the sides of the vial to ensure that no particulate matter is clinging to
the sides of the vial. If solids are observed in the control fluid or on
the vial wall, repeat the mixing procedure. If further mixing does not
homogenizethesample,discardthevialandthawafreshvial
.
3. Open the vial and transfer a drop of solution directly from the vial
into the BNP cartridgeand seal the cartridge. If short-term storage
(<4 hrs) is desired, tightly recap the bottle immediately after all
samplingiscomplete,andstoreat2-8˚C.
4. Insert cartridge into an i-STAT 1 Analyzer.
Target Values and
Ranges
See value assignment sheet posted on the APOC website at www.
abbottpointofcare.com. The value assignment sheet displays target values
andrangesexpectedwhencontrolsandequipmentareperformingproperly.
Shouldresultsfalloutsidetherange,refertotheTroubleshootingsectionthat
followstheProcedureforTestingControls.
Always ensure that the lot number and software revision on the value
assignmentsheetmatchesthelotnumberofthevialinuseandthesoftware
revision in the analyzer.
Targetvaluesarespecictothei-STATSystem.Resultsmaydifferifusedwith
other methods.