14-18 Art: 714376-00M Rev. Date: 31-Jan-12
Documentation of
Results
The results of the thermal probe check are stored in the Central Data Station. If
theCentralDataStationisnotavailable,usetheformincludedinthissection
of the Manual to record the results.
PERFORMING CONTROL TEST ON CARTRIDGE
Procedure for Testing
Controls
InitiatingcontroltestsfromtheQualityTestMenuallowsresultstobestored
inseparatecategoriesforthepurposeofdocumentationandreview.
NOTE:
RNAMedicalControlsmust be tested using this control path and not as a patient sample.
A level surface includes running the handheld in the downloader/recharger.
7. Review results