CHECKING THE THERMAL PROBES IN THE i-STAT ANALYZERS
Verification for
Handheld Analyzers
Verify the thermal probe check for the i-STAT Portable Clinical Analyzer and
i-STAT1Analyzerasfollows:
• ExternalElectronicSimulatorusedroutinelyandresults transmitted
to a Central Data Station Version 4: On the CDS, click on System,
then on Simulator Results. Look under the Delta column. Check
thatthereis a valueequaltoor less than0.1(0.1)listed for each
analyzerinuseinthe last 30 days. A value of “--.--” indicatesthat
the conditions to complete the thermal probe check were not met.
Usetheprocedurebelowtocheckanyanalyzerthatdoesnothavea
numeric value listed.
• ExternalElectronicSimulatorusedroutinelyandresults transmitted
to a Central Data Station Version 5: On the CDS, click on Data
Viewer, then on Simulator. Look under the Probe Delta column.
Checkthatthereisavalueequaltoorlessthan0.1(0.1)listedfor
each analyzer in use in the last 30 days. A value of “--.--” indicates
that the conditions to complete the thermal probe check were not
met. Use theprocedurebelowto checkanyanalyzerthat does not
have a numeric value listed.
• ExternalElectronicSimulatorusedroutinely,resultsnottransmitted
to a Central Data Station: Use the procedure below to check the
thermalprobesoneachanalyzereverysixmonths.
• Internal Electronic Simulator used routinely: Use the procedure
belowtocheckthethermalprobesoneachanalyzertwiceayear.
Procedure for
Handheld Analyzers
Check the thermal probes on the i-STAT Portable Clinical Analyzer and i-STAT 1
Analyzerasfollows:
1. If the analyzer and simulator have been stored separately in areas
wheretheambienttemperaturediffersbymorethan3°C(5°F),allow
thesimulatorandanalyzertostandinthesameplace,outofdrafts,
for 30 minutes before inserting the simulator into the analyzer.
Handle the simulator as little as possible to maintain its thermal
uniformity and stability.
2. Insert the simulator into the analyzer.
3. When results are displayed, the difference between the thermal
probescanbeviewedontheanalyzer’sscreen:
PortableClinicalAnalyzer:whileholdingdowntheDISkey,press
the 1 key.
i-STAT 1 Analyzer: press the period key.
4. Interpretation of the thermal probe check value:
Acceptable:avalueequaltoorlessthan0.1(0.1)
Notacceptable:aFAILmessagewitha“t”QualityCheckCodeor
avaluegreaterthan0.1.Repeattheproceduretoconfirmresults.
Contact your Technical Support representative if the repeat thermal
checkvalueisgreaterthan0.1.
Repeattheprocedure:if“--.--”isdisplayed.Takecaretohandle
the simulator a little as possible. It may help to partially insert the
simulator into the analyzer and let it stand for 15 minutes before
insertingalltheway.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals