22-60 Art: 714384-00D Rev. Date: 07/18/06
Electronic
Simulator Viewer
All Electronic Simulator results, both external and internal, are listed in
chronological order with the newest result at the top of the screen. To view all
simulator results together for each analyzer, click the Serial Number column
header to sort the analyzers by serial number or use the Find... option and
sort for one analyzer. To list in chronological order again, click the Date-Time
header.
To view the actual readings taken during the Electronic Simulator check, click
Record
View Extended Simulator Report... Note that the Simulator ID and
Probe Delta
columns can be viewed in the screen above by scrolling to the right.
Unsent Records
Viewer
This viewer is available only if an external interface installed by i-STAT is enabled
in the Central Data Station Customization function. This data viewer displays
records that have not been sent from the Central Data Station to an external
computer system, such as an LIS. The incorrect information can be corrected
using Edit and the corrected record resent.
Unsent results can be removed from the viewer by highlighting the record,
clicking on Record
Mark Selected Records as Sent, or by pressing the F8
Key
.