Rev. Date: 23-May-12 Art: 716969-00J BNP - 11
Plasma Blood
Sample
Concentration
pg/mL)
Diluted Concentration
(pg/mL)
% Recovery
A 590
B 2764
C 5123
A+B
1570 94%
B+C
3992 101%
A+C
2734 96%
A plasma sample was spiked with BNP to a value of approximately 5000 pg/mL and the concentration was
determined by duplicate measurements with i-STAT BNP test cartridges; the result was found to be within
200 pg/mL of the intended target. This sample was subjected to a series of dilutions with fresh, un-spiked
plasma in order to prepare a range of concentrations. The concentration of each sample/dilution was
calculated based on the measured concentration of the initial solution and the dilutions performed. The
diluted samples were then measured in i-STAT BNP test cartridges (N = 6-10). The procedure was repeated
with a whole blood sample. The results of these experiments are summarized in the following table.
Sample Dilution
Calculated [BNP]
(pg/mL)
Measured [BNP]
(pg/mL)
%Recovery
Plasma 1 52 57 110%
Plasma 2 104 114 110%
Plasma 3 259 265 103%
Plasma 4 518 560 108%
Plasma 5 1036 1002 97%
Plasma 6 2072 2277 110%
Plasma 7 3107 3384 109%
Plasma 8 4143 4222 102%
Whole Blood 1 44 41 93%
Whole Blood 2 88 88 100%
Whole Blood 3 269 287 107%
Whole Blood 4 537 554 103%
Whole Blood 5 725 720 99%
Whole Blood 6 1450 1367 94%
Whole Blood 7 3042 2826 93%
Whole Blood 8 4056 3856 95%
Test Limitations
The frequency of suppressed results is affected by atmospheric pressure. Suppressed result rates may
increase with higher elevations (decreased barometric pressure) and may become persistent if testing is
performed at more than 7500 feet above sea level. Where unavailability of results is unacceptable, i-STAT
recommends having an alternate test method available.
Samples from patients who have been exposed to animals or who have received therapeutic or diagnostic
procedures employing immunoglobulins or reagents derived from immunoglobulins may contain antibodies,
e.g. HAMA or other heterophile antibodies, which may interfere with immunoassays and produce erroneous
results.
28-34
The generation of potentially interfering antibodies in response to bacterial infections has been
reported.
28
While this product contains reagents that minimize the effect of these interferents, and QC
algorithms designed to detect their effects, the possibility of interference causing erroneous results should
be evaluated carefully in cases where there are inconsistencies in the clinical information.

Table of Contents

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals