Horizontal and acquisition setups Autoset
On MSO5000 m odels and DPO5000 models with option MSOE, you can turn on or turn off MagniVu.
What do you want to do next?
Learn about Acquisition modes. (see page 836)
Learn about MagniVu. (see page 159)
Autoset
From the H oriz/Acq menu, select Autoset or push the front panel Autoset button.
Overview
Use Autoset to automatically se t up t he front-panel controls based on the characteristics of the input signal.
This is much faster and easier than manually setting up the controls.
Autoset may change the vertical position to place the waveform appropriately on screen. The instrument
always sets vertical offset to 0 V.
To use
If you use Autoset when one or more channels are displayed, the instrument selects the lowest numbered
channel for horizontal scaling and triggering. The instrument individually scales all displayed vertical
channels. If you use Autoset with no vertical channels displayed, the instrument turns on and scales
channel 1.
The Autoset Undo
(see page 185) button opens in a control window after the instrument completes the
operation. Click this button to undo only the parameters adjusted by the last Autoset.
NOTE. The Autoset Undo control window remains on the screen until you open another control window.
After Au
toset Undo closes, you can still undo the last Autoset by selecting the Undo Last Autoset command
from the Edit menu or the Horiz/Acq menu. Although the last Autoset is immediately undone, the Autoset
Undo control window does not open again.
Select User Preferences from the Utilities menu to display the Prompt Before Action (see page 555)
window.
Click Autoset to toggle between Off and On. Off immediately performs an Autoset when the Autoset
button is pushed. On displays a prompt when the Autoset button is pushed.
Click Close to save your selection.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 183

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals