Trigger setups Set up a LIN bus trigger
To use
Click the Trigger Type box and select Bus from the drop-down list.
Click the Bus box and select the bus number or name from the drop-down list.
NOTE. You have the option to add user-defined label for bus sources.
Click the Logic Thresholds Setup button to set the voltage threshold levels for the channels in the bus.
Select the type of bus cycle or activity to u se as the trigger from the Trigger On drop-down list.
For some Tr
igger On selections, you must set additional fi elds to define other parameters such as for
a Ident & Data.
In this example, you need to set the Trigger When Identifier, When Direction values, and select the And
when data is values. You can also set the component threshold levels through the Logic Thresholds
Setup button.
Sel
ect the Trigger On selections to view their parameters.
What do you want to do next?
Learn about bus setups. (see page 95)
Set up a LIN bus trigger
NOTE. The Bus Trigger Control Window for LIN Serial option is available only on instruments with
option SR-AUTO.
From the Trigger menu, select Bus setup.