Horizontal and acquisition setups FastFrame control windows
Sample mode. It then determines the location of each sample point and accumulates it with shared data
from previous acquisitions. This is the only mode that can be used when doing mask testing with a
optical stand
ard. Fast Acq is inhibited in Waveform Database mode.
What do you want to do next?
Learn about the Horizontal Setup control window. (see page 177)
Learn about the Acquisition Setup control window. (see page 181)
Acquisition interval
The acquisition interval is the time between sample points. For mathematical purposes, it is the time
covered by the waveform record divided by the record length.
FastFrame control windows
From the Horiz /Acq menu, select FastFrame Setup.
Overview
Use the FastFrame control windows to capture and view a series of triggered acquisitions without wasting
acquisition memory on the dead time
(see page 187) between the trigger events.
For information o n the controls, click the buttons.
Behavior
FastFrame mode combines multiple acquisitions into one large data record. By carefully defining the
trigger conditions, you can us e FastFrame to capture o nly multiple events
(see page 196) of the w aveform
that you are interested in. Each captured event is stored sequentially in its own frame
(see page 879).
The frames are in the order they were captured.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 195