Mask testing setups Set up limit testing
Behavior
The Recall dialog box displays when you click Recall in the Mask Edit control w indow. Clicking Recall
replaces the current mask with the mask that you selected in the Recall dialog box.
Masks are saved in
C:\Users\[Username]\Tektron ix\TekScope\Mas ks unless you choose another
directory.
What do you want to do next?
Learn about saving masks. (see page 536)
Learn more about mask types. (see page 209)
Learn more
about mask setups.
(see page 209)
Set up limit testing
From the Mask menu, select Limit Test Setup; then open the Setup tab.
Overvi
ew
Limit testing allows you to compare an active signal to a template waveform.
To use
Use the following procedure to build a template from a known signal and compare it to an active signal.