EasyManuals Logo
Home>Tektronix>Test Equipment>DPO5000 Series

Tektronix DPO5000 Series User Manual

Tektronix DPO5000 Series
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #797 background imageLoading...
Page #797 background image
Oscilloscope Reference Gaussian window
What do you want to do next?
Learn about using spectral analyzer windows. (see page 771)
Gaussian window
This is the default window function (see the next gure). It is unique in that the time-domain shape o f an
exponential Gaussian function transforms into a Gaussian exponential shape in the frequency domain.
NOTE. This window provides optimal localization in both the time and the frequency domain
What do you want to do next?
Lea
rn about using spectral analyzer windows.
(see page 771)
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 775

Table of Contents

Other manuals for Tektronix DPO5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO5000 Series and is the answer not in the manual?

Tektronix DPO5000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals