Vertical Setups Termination Controls
Using Offset to Avoid Clipping Signals
Clipping
The circled portion of the incoming waveform is clipped,
and therefore, is not acquired.
The clipped portion cannot be processed or displayed.
Offset can move the acquisition window vertically relative
to the incoming waveform.
Offset removes clipping
By setting an offset to vertically shift the acquisition
positive relative to the waveform, the clipped portion can
be acquired.
xxx
Termination Controls
From the Vertical menu, select Termination.
Overview
Use the Termination control window to select the channel input resistance.
For 4 GHz model oscilloscopes, Termination is xed at 50 .
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 489