6-12 CASE Revision C
2060290-201
Before the Test
Operating Steps
1. Check the applied electrodes and verify the signal quality in all
leads.
ST measurements and arrhythmia analysis will automatically be
displayed when all necessary electrodes have been applied and
signals of adequate quality are obtained (e.g. without major baseline
fluctuations). If all electrodes are not being used, click Relearn
Medians to manually activate ST measurement and arrhythmia
analysis.
2. Click Select Protocol to verify that the appropriate test protocol is
active or to select another protocol.
3. Click Patient Education to display the patient exercise test
instructions.
4. Click Custom Setup to select and load user-specific test settings.
5. Click Setup to edit the exercise test setup.
Note
Once you initiate the pretest phase you can no longer change the
test settings.
6. Press pretest to initiate the test. The pretest screen will appear and
the pretest phase begins.