Revision C CASE 7-5
2060290-201
Setting Up the Ambulatory Blood Pressure Monitor
Operating Steps
1. Turn the monitor off and on again.
2. Wait for the time to be displayed, then connect the monitor to the PC.
3. Click Connect.
The acquisition screen indicates the memory status of the ambulatory
blood pressure monitor.
A: Date and time of the monitor.
B: Measurement protocol.
C: Number of BP measurements and the duration of the protocol.
D: Information whether or not data is stored in the monitor.
With TONOPORT V, version 2.0 and later:
E: Patient ID stored in the TONOPORT device.
F: Patient data from the CASE database assigned to the patient ID.
4. Click the Clear Data button to delete data stored in the monitor (with
TONOPORT V devices, version 2.0 and later, this command will also
clear the patient ID).
Note
If the stored data has not been transferred to the system,
download it as described in the next section.
5. Click Programming to select the protocol.
182A-BP
ABC
DE F