5-10 CASE Revision C
2060290-201
Recording a Resting ECG
Full Disclosure ECG
You can acquire the full disclosure ECG over a maximum period of 60
minutes.
Operating Steps
1. Click Start Full Disclosure ECG to initiate ECG storage.
2. While the full disclosure ECG is acquired:
a. You can make an ECG recording (Rhythm Writer, Stop Writer).
b. You can click Event Marker Full Disclosure to enter event marks
(vertical line).
c. The most recent arrhythmia is always displayed.
d. The remaining storage time, the total number of stored QRS
complexes and the number or arrhythmic events is displayed.
3. Click Stop FD/Post Test Review to terminate the procedure.
Note
Clicking Stop FD/Post Test Review brings up the Test Summary
(see page 5-11).
056A-R
2c
2a
2b
1
3
2d