6-42 CASE Revision C
2060290-201
After the Test — Post Test Review
Performing Reanalysis
1. Click a measurement mark above the baseline and hold the mouse
button depressed.
2. Move the mark to its new position and release the mouse button.
3. Adjust the other marks in the same way.
4. Select the complexes to reanalyze.
5. Click the Reanalyze button.
6. Click ...Reset and Reanalyze... to restore the original analysis results.
(Reset to original restores the original results for all complexes of the
entire exercise test).
Performing Measurements on the Median Complex
Amplitudes (mV), intervals (ms) and slopes (mV/s) can be measured on
all sample cardiac cycles:
1. Position the cursor on the first measurement point and click. A
caliper appears.
2. Position the cursor on the second measurement point and click. Now
you will see
the second caliper
the measuring line
the measurement results
3. Click the right mouse button to remove the measurement points.