EasyManuals Logo
Home>GE>Medical Equipment>CASE V6.7

GE CASE V6.7 User Manual

GE CASE V6.7
342 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
Revision C CASE 6-43
2060290-201
After the Test — Post Test Review
Sample TWA Cycles
Overview
This screen shows the Sample TWA Cycles. Questionable values are
identified with a question mark ?. Two-headed arrows identify the
position where the TWA value was determined. Individual leads can be
excluded. You may then repeat the analysis without these leads. This
method prevents false-positive TWA values, particularly when
permanently disturbed leads are excluded.
938A-S
B
C
D
E
F
H
A
G
A Click to view more sample TWA cycles.
B Click to show and hide the measurement marks.
C Select the speed of the displayed medians.
D Select the gain for the displayed medians.
E Select the number of lines (1, 3, 6).
F Number of medians shown per stage:
: All cardiac cycles (at intervals of 30 seconds)
: only one cardiac cycle
G Position where the TWA value was determined.
H Click to exclude individual leads.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CASE V6.7 and is the answer not in the manual?

GE CASE V6.7 Specifications

General IconGeneral
BrandGE
ModelCASE V6.7
CategoryMedical Equipment
LanguageEnglish

Related product manuals