GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
4-26 Section 4-2 - General Procedure
4-3-10-3 Caster and Caster Link Checks
Check if:
•For Caster:
- No loose caster with it jacked up.
- Smooth operation when moving the scanner.
For Caster Link Assy:
- Effective brake and direction lock function (See section 4-3-10-2).
- The pedals can be smoothly returned.
- The lock pin must be removed.