DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 6-7 - Cleaning the Trackball 6-35
Section 6-7 Cleaning the Trackball (cont’d)
7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud.
NOTICE
When cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard
or system). Use either ethanol, isopropyl alcohol or VCR head cleaner to clean the trackball assembly.
Avoid other solvents that may damage the mechanical parts of the trackball assembly.
Figure 6-44 Cleaning Rollers
Rollers
Cotton Bud