DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
6-10 Section 6-5 - Monitor and LCD Adjustments
Section 6-5
Monitor and LCD Adjustments
6-5-1 CRT Monitor Contrast and Brightness Adjustment
Before calibrate the Monitor, check your PCVIC or DGVIC Part Number on the Utility Screen. The typical
setting value changes with its Part Number.
(To enter this screen, select Utility > System > About > Additional About Information.)
To adjust the contrast and brightness:
1.) Select Calibration from the Test Pattern Utility Touch Panel. The test pattern consists of a small
box inside a larger box.
Figure 6-10 PCVIC/DGVIC Part Number Check
Figure 6-11 Calibration