GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
2-10 Section 2-3 - Facility Needs
2-3-5 Networking Pre-installation Requirements
2-3-5-1 Purpose of DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-5-2 DICOM Option Pre-installation Requirements
To configure the LOGIQ™ 7 to work with other network connections, the site’s network administrator
must provide some necessary information.
Information must include:
A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ™ 7.
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ™ 7 for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and
the revision of the device, is also included. This information may be useful for solving errors.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals