GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 6-2 - LV Unit Adjustments (For BT03 or lower) 6-3
6-2-2 Adjustments Procedures
1.) Using the following VRs, DC output can be adjusted.
Table 6-2 DC Output Specification for LV Unit
VR
Specifications
VR1
3.4V +/- 34mV
VR2
5V +/- 50mV
VR3
6V +/- 60mV
VR4
-6 V +/- 60mV
VR5
12V +/- 120mV
VR6
-12V +/- 120mV
Figure 6-2 VRs for Adjusting DC Output
VR4 for -6V
VR2 for +5V
VR3 for +6V
VR6 for -12V
VR1 for +3.4V
VR5 for +12V
OR
-6V
+12V
+5V
-12V
+3.3V
+6V