DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
4-8 Section 4-2 - General Procedure
4-2-4 System Stand-by (cont’d)
7.) To power up the scanner, tap POWER button momentary, and wait for power up. It will
approximately 45seconds for the scanner to power up.
Stand-by Sequence
NOTICE
Do not press the Power button long time, as it will cause the system to power down.
Figure 4-11 Stand-by Sequence
Press POWER button
Select "Standby"
Check Monitor
OFF
Wait
Check Touch
Panel OFF
Check POWER
button
illuminated
Wait more than 20 seconds
Disconnect Power Cord
Connect to electrical outlet
(Power Cord)
Press POWER button
Wait 5 seconds
Scanner Start up completion
in 45 seconds
Standby may be
failed
Stand-by
Se
q
uence
ON
ON
OFF
OFF or Blinking
(Note 1)
Note1 : BT09 Release with BEP4 Hardware have "Blinking Power Button"
capability to indicate the console in Stand-by condition. For BEP3 and earlier,
Power button must be "OFF" before disconnecting the power cord.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals