6-30 Section 6-6 - BW Printer Setting / Adjustment
6-6-2 Parameters for UP-D895 (cont’d)
6.) Click on Scaling..., then check off into "Enlarge to Paper" and set "Bilinear" into "Interpolation
Method".
7.) Click on Density Adjust..., set "TONE1" into "GAMMA" and set 7 into "Sharpness".
6-6-3 Service Tips - Print Quality
When printer is set to “Nearest Neighbor” interepolation method, comments and body pattern may
become hard to read. Re-adjust interpolation method to “Bi-Linear” to avoid this situation.
Figure 6-39 Scaling
Figure 6-40 Density Adjust