Section 6-5 - Monitor and LCD Adjustments 6-23
6-5-5-2 For Technart (cont’d)
e.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
f.) Repeat to press the MENU button until the “Phase” blinks.
g.) Repeat to press the PLUS button to find the value with no character blurring. For example, this
value is X.
h.) Set the value (X + 14) to Phase.
i.) Press the AUTO button once to save the setting.
j.) Repeat to press the MENU button until the “Exit” blinks
k.) Press the AUTO button once to exit from the menu.
Figure 6-30 Menu Screen
NOTICE
You can select the number from 0-62. Values with deblurring cause are continuous seven or eight
values. So, tips for searching the best value is to avoid such a group. This depends on video card or
VGA cable of your scanner.