GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-39
8-2-10-6 VCR Parameters Setting (for the VCR-equipped system ONLY)
1.) Touch Utility > System.
2.) Click on Peripherals tab.
3.) Set the proper parameters:
-PAL or NTSC
- VCR type
4.) Shut down the scanner, then restart the scanner.
5.) Verify that the VCR can be properly operated with no error (remote, play...) .
8-2-10-7 Regional Setup (For the system with language other than English ONLY)
For Japanese Languages
You can set up the keyboard for Japanese. If you select Japanese, the following items display or enter
by Japanese.
Patient name
Exam comment in Patient screen and Worksheet
Report/Direct Report free text, Insert Text
Annotation, Annotation Library
Error/Warning message
DICOM/Worklist, Query/Retrieve, Image Storage, Printer, MPPS
Figure 8-99 Peripherals