GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
2-6 Section 2-2 - General Console Requirements
2-2-4 Probes Environmental Requirements
Table 2-6 Operation and storage Temperatures for 2D Probes
Operational Storage Transport
Temperature
10 - 40
o
C
50 - 104
o
F
-10 - 60
o
C
14 - 140
o
F
-40 - 60
o
C
-40 - 140
o
F
Humidity
30 - 85%
non-condensing
30 - 90%
non-condensing
30 - 90%
non-condensing
Pressure
700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
Table 2-7 Operation and storage Temperatures for 4D Probes
Operational Storage Transport
Temperature
18 - 30
o
C
64- 86
o
F
-10 - 50
o
C
14 - 122
o
F
-10 - 50
o
C
14 - 122
o
F
Humidity
Max. 70%
non-condensing
Max. 90%
non-condensing
Max. 90%
non-condensing
Pressure
700 - 1060hPa 700 - 1060hPa 700 - 1060hPa