DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-51
8-2-10-10 Regional Preset
9.) Touch Utility > System.
10.)Click About tab.
11.)Click Additional About Information.
12.)Check the region name of Preset Region (Rquires reboot):
If it is None, skip this section.
13.)Click Reload Factory Default button.
14.)Click OK on confirmation dialog box.
This changes Preset Region (Requires reboot) to None.
NOTICE
For the system other than "NONE" selected as Regional Preset, resetting MUST be required.
Figure 8-113 Preset Region
Figure 8-114 Reload Factory Default