Roche Diagnostics
Operatorโs Manual ยท Version 3.1 A-5
Cedex Bio System 1Safety
Safety classification
Safety classification
Before you attempt to use the Cedex Bio instrument, you must be fully familiar with
the following symbols and their meanings:
Safety information
Before operating the Cedex Bio instrument, it is essential that you both read and
understand the safety information listed below.
Read all Roche safety notices carefully and make sure you understand them.
Transport
Electrical safety
Warning
Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
Caution
Indicates a hazardous situation which, if not avoided, may result in minor or moderate
injury.
Notice
Indicates a hazardous situation which, if not avoided, may result in property damage.
Injury from heavy loads
You may injure your hands, fingers, or back when putting the analyzer in place. Carry the
analyzer according to the transport instructions.
Electrical shock by electronic equipment
Do not attempt to work in any electronic compartment. Installation, service, and repair
must only be performed by authorized and qualified personnel.
Electrical safety
Connect the analyzer to grounded power outlets only (IEC protection class 1). All periph-
eral devices that are connected to the Cedex Bio instrument must comply with safety
standard IEC 60950 for information technology equipment, or with IEC 61010-1, UL 61010-
1 for laboratory use instruments.