Roche Diagnostics
B-154 Operatorโs Manual ยท Version 3.1
7 Configuration Cedex Bio System
Applications
Control definitions
Item Values Comment
Sequence No Interval The system does not inform you about the due
status.
Use this value if you intend to perform QC as
part of standard sample testing and not as a sep-
arate task, and if QC should be handled by a host
computer.
Note that an accepted QC result is required for a
test before it can be used.
Interval Only Due whenever the reagent interval has expired.
The system automatically informs you about
due QCs.
Interval Number of days
(hours for ISE
applications)
Use 0 (zero) if the Sequence is set to No Interval.
For each of the controls:
Mat. Code 0 ... 999 Unique identification of the control material.
Numbers 801 through 999 are reserved for non
Roche controls.
You need to define this initially for every appli-
cation.
QC After Cal On, Off QC needs to be performed after each calibration
of the test.
Default QC On, Off Mark the test to have QC performed as part of
the Default QC function.
Default QC is an automated process for per-
forming all QC measurements that are currently
due. This is the ideal method if you want to per-
form QC periodically during routine operation.
Table B-14 Control definitions