Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-61
Cedex Bio System 5 Daily operation
Performing calibrations
Process of performing calibrations
The process of defining calibration orders depends on what you want to achieve:
o Calibrating all tests that need calibrating
o Calibrating all tests that will need calibrating during the forecast period
o Calibrating individual tests
O Calibrating the current set
O Pre-calibrating a standby set
e
For information on performing calibrations in the Prepare phase, see Performing the
calibrations (Prepare phase) on page B-30.
a To perform calibrations
1
Choose Workplace > Calibrations.
A screen is displayed that lists all tests and their currently valid calibration.
2
Press .
A screen is displayed that shows all active tests.
Interpreting the colors
A Press to select all tests that now need
calibrating.
B Press to select all tests that need
calibration within the forecast period.
Figure B-20
A
B
Color Meaning Possible actions
Calibration is due. Perform the calibration.
Calibration failed. Check the result flag to find out why
the calibration failed. Repeat the cali-
bration if necessary.
The calibration was successfully per-
formed, but its result has not been
accepted yet.
Validate the calibration results.
Not relevant for calibration.
The calibration is OK. Its result has
been accepted.
No action is currently required.
The calibration is OK. It applies to a
reagent set that is not active.
No action is currently required.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals