Safety information
Power interruption
Electromagnetic devices
Approved parts
Third-party software
Data loss or damage to the system due to voltage drop
By a power failure or momentary voltage drop the operation unit or software of this system
may get damaged or data loss may occur. Use only uninterruptible power supply.
Malfunction of instrument and incorrect results due to interfering electromagnetic
fields
Devices that emit electromagnetic waves may cause the instrument to malfunction. Do not
operate the following devices in the same room where the system is installed:
o Mobile phone
o Transceiver
o Cordless phone
o Other electrical devices that generate electromagnetic waves
Class B FCC rule compliance
This equipment has been tested and found to comply with the limits for Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reason-
able protection against harmful interferences when the equipment is operated in a resi-
dential area. However, this equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the present manual, may cause
harmful interference to radio communications.
The electromagnetic environment should be evaluated prior to operation of the device.
Malfunction of instrument and incorrect results due to nonapproved parts
Use of nonapproved parts or devices may result in malfunction of the instrument and may
render the warranty null and void. Only use parts and devices approved by Roche Diagnos-
tics.
Malfunction of instrument and incorrect results due to third-party software
Installation of any third-party software that is not approved by Roche Diagnostics may
result in incorrect behavior of the system. Do not install any nonapproved software.