Cedex Bio System 13 ISE operation
Daily operation
Starting the Prepare wizard
a To start the preparation process
1
Choose Overview > Prepare.
Checking the status of the external fluid containers
a To check the status of the external fluid containers
1
Follow the instructions given in Checking the status of the external fluid containers
on page B-15.
2
When you have finished with the fluid containers, press to proceed to the
next stage in the Prepare wizard.
A screen is displayed that shows the status of the electrodes and the on-board ISE
fluids.
Checking the ISE status
Replacing electrodes
a To replace an electrode
1
Follow the instructions given in Replacing electrodes on page E-32.
When checking external fluid containers without using the Prepare wizard:
Choose Overview > > .
Injury through working solutions
Direct contact with cleaning solutions or other working solutions may cause personal
injury.
When handling such solutions, exercise the precautions required for handling them,
observe the cautions given in the package insert, and observe the information given in the
Safety Data Sheets available for Roche Diagnostics cleaning solutions.
o After replacing an electrode, the Electrode Service maintenance action needs to
be performed. This action will be performed as part of the maintenance procedure
of the Prepare wizard.
o After replacing an electrode, a main calibration needs to be performed. This action
will be performed as part of the Daily Prepare Actions maintenance action.