Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 C-23
Cedex Bio System 8 General maintenance
Maintenance actions
Clean up the database
f Utilities > Maintenance > Cleanup Database
This maintenance action restores the database to the condition where the instrument
can work at its optimum. You typically use this action when you think that the
instrument suddenly works slow, for example when displaying information on the
screen takes long or when it takes long until an action starts after you have initiated it.
Operator time Approximately 10 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required None
a To clean up the database
1
Select the maintenance action Cleanup Database.
2
Press .
The maintenance definition screen is displayed.
3
Press .
An instructions screen is displayed while the database cleanup is performed.
4
When the maintenance action is complete, press to confirm the completion.