Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 B-81
Cedex Bio System 5 Daily operation
Finishing the shift
Checking the transmission of results
This step is only relevant if you work with a host system.
Before deleting the results, you must make sure they were properly transmitted to the
host.
a To check for results that were not transmitted
1
Choose Workplace > Result Review.
A screen is displayed that contains all sample results of the day.
2
Choose > Not Sent to Host.
3
Press .
A screen is displayed that contains the sample results that have not been
successfully transmitted to the host.
4
Select a result.
5
Press .
A screen for validating results is displayed.
6
Press Retransmit.
The list with the results that were not transmitted successfully is displayed again.
The result you just retransmitted should no longer be present.
7
Select and retransmit the remaining results as described in Steps 4 through 6.
8
Choose > Not sent to Host.
The list should now be empty.
9
If the results were not successfully transmitted, contact the administrator of the
host computer.
Starting the End Shift wizard
The easiest way of performing the end of shift tasks is by following the End Shift
wizard.
a To start the End Shift wizard
1
Choose Overview > End Shift.
A screen is displayed for performing daily backup.
Results must be accepted before they can be transmitted to the host.
See also the information on Result handling on page B-165, in particular on the Auto
Accept settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals