EasyManuals Logo
Home>Roche>Measuring Instruments>Cedex Bio System

Roche Cedex Bio System User Manual

Roche Cedex Bio System
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #441 background imageLoading...
Page #441 background image
Roche Diagnostics
Operatorโ€™s Manual ยท Version 3.1 E-51
Cedex Bio System 14 ISE maintenance
ISE maintenance actions
Clean ISE tower manually
f Utilities > Maintenance > Clean ISE Tower Manually
Operator time Approximately 6 minutes.
Prerequisites The system must be in Standby status.
Tools and materials required m ISE Deproteinizer
m Activator
m Cotton swabs
a To clean the ISE tower
1
Select the maintenance action Clean ISE Tower Manually.
2
Press .
The maintenance definition screen is displayed.
3
Press to start the action.
A placement list is displayed.
4
Place the Activator and ISE Deproteinizer on the sample area positions indicated
on the screen.
5
Press to confirm the placement and to start the action.
The ISE tower is first washed.
The transfer head moves to the right side of the instrument.
You are asked to clean the ISE tower.
6
Open the main cover and then the left service flap.
When the ISE tower is blocked, perform first the Clean ISE Tower Automatically mainte-
nance action (See Clean ISE tower automatically on page E-50). If this does not help, per-
form the Clean ISE Tower Manually maintenance action. If the ISE tower is still blocked,
clean the ISE tower off the instrument. (See Cleaning the ISE tower off the instrument on
page E-36.)
Make sure that you have read and understood section Safety information on page E-42.
The following warning messages in particular are relevant:
o Injury through working solutions on page E-42.
o Infection by biohazardous materials on page E-42.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals