9 Messages and alarms Cedex Bio System
List of alarm messages
List of alarm messages
In the following table, the variables %s and %d are used, they represent text and numbers.
ID Message Comment
0000 Unknown message, id=%d!
0001 A fatal system exception software error (%d) occurred. Restart the system. If the error recurs export the log files
while the system stays in the error condition. If the error
persists call service.
0002 An internal software error (%d) occurred while updating
the configuration.
Restart the system. If the error recurs export the log files
while the system stays in the error condition. If the error
persists call service.
0003 A software error occurred. Order (%d, %d) could not be
processed.
Delete the orders that are not processed yet. If the error
recurs export the log files while the system stays in the
error condition. If the error persists call service.
0004 A software error occurred. The unit is not ready, order
(%d) rejected.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0005 A mathematical exception arose. Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0006 Last calculation cycle (%d) too short. Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0007 Failed to calculate linear regression. Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0008 The cycle range (%d..%d) is not valid. Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0010 The reaction direction (%d) is invalid. Repeat the order. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0011 An internal communication error occurred. The received
IPC message (%d, %d) could not be interpreted.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
0012 A software error occurred. The status of an order is
invalid (%d.%d).
Delete and redefine the order. If the error recurs export
the log files with the system remaining in this error condi-
tion. If the error persists call service.
0017 The installed language version is out of date (%s, %s). Install the language version that belongs to the currently
installed system software version. If the error persists call
service.
0018 The current system date and time are invalid. Correct the date and time settings in Configuration. If the
error persists call service.
0019 Printer controller revision %d is not supported. Call service.
Table D-1 List of alarm messages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals