Roche Diagnostics
E-24 Operatorโ€™s Manual ยท Version 3.1
13 ISE operation Cedex Bio System
Daily operation
Monitoring the progress
a To monitor the progress
1
Follow the instructions given in Monitoring the analysis progress on page B-44.
a To check the instrument status
1
Press the Overview tab.
Interpreting the instrument
status buttons
A Relevant instrument status buttons
Figure E-4
A
Icon Meaning Possible action
The ISE unit is in standby. No action is required.
The expiration date of an electrode
has passed or an on-board ISE
fluid level is low.
1. Check the details on the dedi-
cated System Status screen.
2. Replace the electrode or ISE
fluid bottle as soon as possible.
The ISE unit is blocked. 1. Check the details on the dedi-
cated ISE status screen.
2. Act accordingly.