Cedex Bio System 8 General maintenance
Maintenance actions
Contact service representative
f Utilities > Maintenance > Call Service Rep.
To ensure smooth operation of the instrument, a service representative periodically
needs to perform some preventive maintenance tasks.
The system displays a message when this maintenance action is due.
Operator time Approximately 2 minutes.
(A service representative takes about 4 hours to perform preventive maintenance.)
Prerequisites None
a To initiate preventive maintenance
1
A message screen reminds you that you should contact your service
representative.
2
Select the maintenance action Call Service Rep.
3
Press .
The maintenance definition screen is displayed.
4
Press .
An instructions screen is displayed.
5
Contact your service representative to arrange for preventive maintenance.
6
Press .