Roche Diagnostics
B-80 Operatorโ€™s Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Finishing the shift
Checking for unfinished tasks
Checking for unfinished orders
a To check for unfinished orders
1
Choose Workplace > Orders.
2
Choose > Not Finished.
3
Press .
A screen is displayed that contains all unfinished orders.
4
Do whatever is necessary to finish the orders.
Checking for not accepted sample results
a To check for not accepted results
1
Choose Workplace > Result Review.
A screen is displayed that contains all sample results of the day.
2
Choose > Not Accepted.
3
Press .
A screen is displayed that contains all sample results that have not been accepted
yet.
4
Select a result.
5
Press to validate the result.
A screen is displayed for selecting your decision.
e
See Accepting results on page B-57.
See Repeating tests on page B-55.
See Rerunning tests on page B-56.
6
Press one of the buttons.
7
Select and validate the remaining results as described in Steps 4 through 6.
If you did not accept all results, deal with the not accepted ones and check for not
accepted results again.
Results must be accepted before they can be printed or transmitted to the host.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals