Roche Diagnostics
B-72 Operatorโs Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Performing QC
5
Press a control button.
A screen is displayed, asking you to place the selected control.
6
Place the control tube on any free position on the sample area.
The system registers the position and automatically defines an order.
If there are more controls to place, the screen with the control buttons is displayed
again. The control you just placed is no longer active.
7
Select and place the remaining controls as described in steps 5 through 6.
When all controls have been selected and placed, the Overview tab is displayed.
8
Press to start processing the control orders.
Validating QC results
Flagged QC results If flags are generated, you must determine their cause and decide whether to accept
the result or to ignore it.
The effect of flagged QC results depends on the configuration.
e
See Result handling on page B-165.
See Editing the acceptable flags list on page B-160.
Configuration setting Effect
QC Auto Accept: On Results that do not contain a flag are auto-
matically accepted. Results with flags that are
marked in a predefined list of flags that
should be ignored are also automatically
accepted.
QC Auto Accept: Off All results need to be manually accepted.
QC Acceptable Flags Flags that are marked in this list will be
ignored by the system.
Table B-8 QC flag configuration and its effect