Roche Diagnostics
B-64 Operatorโ€™s Manual ยท Version 3.1
5 Daily operation Cedex Bio System
Performing calibrations
Validating calibration results
For a calibration to become active, you need to accept its result.
Flagged calibration results If flags are generated, you must determine their cause and decide whether to accept
the result, continue working with the old results, or to rerun the calibration.
The effect of flagged calibration results depends on the configuration.
e
See Result handling on page B-165.
See Editing the acceptable flags list on page B-160.
Configuration setting Effect
Cal Auto Accept: On Results that do not contain a flag are auto-
matically accepted. Results with flags that are
marked in a predefined list of flags that
should be ignored are also automatically
accepted.
Cal Auto Accept: Off All results need to be manually accepted.
Cal Acceptable Flags Flags that are marked in this list will be
ignored by the system.
Table B-6 Calibration flag configuration and its effect

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Cedex Bio System and is the answer not in the manual?

Roche Cedex Bio System Specifications

General IconGeneral
BrandRoche
ModelCedex Bio System
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals