Cedex Bio System 6 Special operations
Connecting and disconnecting the external fluid containers
6
Install the transfer head cover.
Push down until the release buttons latch on.
7
Switch on the system.
The system performs internal checks and routines.
The startup phase may take a few minutes.
When the instrument is ready for use, the Overview tab is displayed, and its status
is Standby.
8
Log on the system.
9
Perform the Deproteinize Probe maintenance action.
e
See Deproteinize the probe on page C-10.
Connecting and disconnecting the external fluid containers
Before switching on the system, make sure that all external fluid containers are placed
on the rack and properly connected.
a To connect the water container
1
Place the white full water container on the tray. (Always use purified water of the
quality specified in section Technical specifications.)
2
Connect the water tubing to the connector on the instrument by pushing the plug
firmly in the socket.
3
Insert the tubing adapter in the container and push it down firmly.
A Water container
B Tubing adapter
C Water connector
Figure B-41