GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
3-64 Section 3-13 - Configuring Connectivity
3-13-6-3 Changing parameters for a service
There are certain parameters that may need to be set up for each service:
Many service parameters are specific to each type of service. The parameters are described on the
following pages:
DICOM Image Storage
DICOM Performed Procedure
DICOM Print
DICOM Query/Retrieve
DICOM Storage Commitment
DICOM Worklist
Standard Print
Video Capture
Save As
Network storage
USB Quick Save
Table 3-26 Service Parameters: Common Service Parameters
Preset Parameter Description
Name Free text: give a descriptive name to the device.
AE Title The Application Entity Title for the service.
Port Number The port number of the service.
Maximum Retries Max # - the maximum number of times to try establishing a connection to the service.
Retry Interval (sec) Specify how often (in seconds) the system should try to establish a connection to the service.
Timeout The amount of time after which the system will stop trying to establish a connection to the
service.

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals