GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-7
Section 8-3
System Software - Installation/Upgrade Procedure
8-3-1 Introduction
This section describes procedure to install/upgrade software to system.
8-3-2 Manpower
1 person, 1 hour
- Depends on contents of System DVD, peripherals, etc. (R1.x.x or R2.x.x)
- Depends on contents of USB media, peripherals, etc. (R3.x.x or later)
8-3-3 Tools
- System DVD (R1.x.x or R2.x.x)
- Software Flash Drive with Base Load Image and Application Software. (R3.x.x or later)
8-3-4 Preparations
Before performing the Software Upgrade:
make sure that all system functions are working correct
check the current Application Software version and the installed Options
8-3-5 Preparation
8-3-5-1 General Notes on LOGIS P7/P9 Software Loading
When preparing to load software, you must perform the following steps:
Back up the entire Patient Archive and Reports (you can import the database from previous version
into this version).
Back up all presets, including Service.
Record Connectivity settings (TCP/IP).
Load Base System Software and install Application Software:
- Reload entire drive (all partitions) or
- Reload only C partition

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals