EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P7

GE LOGIQ P7 Service Manual

GE LOGIQ P7
602 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #450 background imageLoading...
Page #450 background image
GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-124 Section 8-22 - Replacement of Caster
Section 8-22
Replacement of Caster
8-22-1 Manpower
2 persons, 15 minutes.
8-22-2 Tools
24mm Socket or adjustable Wrench, Standard Phillips Screwdriver
NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
8-22-3 Removal Procedure
Figure 8-147 Caster
NOTE: Press the caster wheel locks and ensure that they prevent the system from moving.
1) Remove Front Cover (refer to section 8-6)
2) Remove Top Cover (refer to section 8-7)
3) Remove Side Cover (refer to section 8-8)
Total Lock
Total Lock
Total Lock
Swivel Lock

Table of Contents