GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
8-174 Section 8-44 - Replacement of MCWD ASSY
Section 8-44
Replacement of MCWD ASSY
8-44-1 Manpower
1 person, 15 minutes
8-44-2 Tools
Standard Phillips Screwdriver.
8-44-3 Removal Procedure
1) Separate MCB Assy and MDC Assy. (Refer to section 8-17-7)
2) Unscrew 1 screw and remove MCWD assy
NOTE: Be careful not to damage MCWD connector.
Figure 8-205 Removal of MCWD board.
8-44-4 Installation Procedure
Parts to be installed in reverse order on removal.
8-44-5 Functional Check
NOTE: Refer to section 4-4 Functional Diag Test List