EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P7

GE LOGIQ P7 Service Manual

GE LOGIQ P7
602 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #318 background imageLoading...
Page #318 background image
GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
7-24 Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7-7-4 Printer Malfunction
Figure 7-19 Printer Malfunction
Printer malfunction
Is printer
properly
connected?
Check all signal and power supply cable
connections between Printer and Scanner
Is printer
loaded with paper?
Load paper by using the correct format.
Check paper size (small, large, etc.)
Check paper tray to fit with the used paper format
Is printing media
(ink, cartridge, etc.)
empty?
Depending on the Printer Type
replace ink, cartridge etc. color
Are the P-keys
configured properly?
Check programmed P-keys and
re-configure them if not correct
Is printing also
impossible by manual
operation?
Check remote cable
Perform Printer Test
Print Test page
If unsuccessful replace printer
Return to normal scanning
yes
yes
yes
yes
yes
yes
no
no
no
no
no

Table of Contents

Other manuals for GE LOGIQ P7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P7 and is the answer not in the manual?

GE LOGIQ P7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P7
CategoryMedical Equipment
LanguageEnglish

Related product manuals