GERAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
7-10 Section 7-4 - Screen Captures and Logs
7-4-3-3 System voltage failure
7-4-3-3-1 Cause
System detected unexpected voltage level from voltage level monitoring circuits.
7-4-3-3-2 What to Do
1) Press OK and reboot the system.
2) If problem persists, then shut down the system, turn OFF breaker and reboot.
7-4-3-3-3 Log message of the Voltage error
(Example)
Error ; acqHouseKeeping(540); MCB_AFE_3V3_0= 2.5 outside tolerance limits. UpperBound =
3.6 LowerBound = 2.970, Soft Limits are usually narrower than upperBound and lower bound