GEDRAFT LOGIQ P9/P7
D
IRECTION 5604324, REVISION 11 DRAFT (JANUARY 24, 2019) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-9
7-4-3 System Warning/Errors and Logs
7-4-3-1 Temperature warning
7-4-3-1-1 Cause
CPU Temperature exceeded threshold (104°C).
7-4-3-1-2 What to do
1) Shut down the system and Clean the filter.
7-4-3-2 Temperature exceeded threshold
7-4-3-2-1 Cause
CPU Temperature exceeded threshold (100°C) and therefore, system must be shut down immediately.
7-4-3-2-2 What to do
1) Shut down the system and Clean the filter.
7-4-3-2-3 Log message of the high temperature
(Example)
Error ; acqHouseKeeping(****); CPU Temperature Sensor BEP_CPU_TEMPERATURE = 105.00
degrees C is outside tolerance limits. UpperLimit = 104.00 LowerLimit = -10.00